• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX TPE; SEPARATOR FOR THERAPEUTIC PURPOSES, MEMBRANE AUTOMATED BLOOD CELL/PLASMA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MEYZIEU PRISMAFLEX TPE; SEPARATOR FOR THERAPEUTIC PURPOSES, MEMBRANE AUTOMATED BLOOD CELL/PLASMA Back to Search Results
Catalog Number 115315
Device Problem Misassembled (1398)
Patient Problem No Patient Involvement (2645)
Event Date 04/14/2018
Event Type  malfunction  
Manufacturer Narrative
Facility name - the first hospital affiliated to (b)(6) medical university.The actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection of the provided photograph shows that the pbp and effluent lines are incorrectly assembled on the support plate.The reported condition was verified.The cause of the condition was determined to be an operator error during manufacturing.To address this issue, awareness training was provided to all operators.The reported lot was manufactured prior to this action.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the effluent line (yellow) was mismatched with the pre blood line (white) of a prismaflex tpe2000 set ckt.This was identified during priming.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX TPE
Type of Device
SEPARATOR FOR THERAPEUTIC PURPOSES, MEMBRANE AUTOMATED BLOOD CELL/PLASMA
Manufacturer (Section D)
BAXTER HEALTHCARE - MEYZIEU
meyzieu cedex rhone
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
7, av lionel terray, b.p. 126
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7500757
MDR Text Key108248811
Report Number8010182-2018-00039
Device Sequence Number1
Product Code MDP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P830063
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/01/2019
Device Catalogue Number115315
Device Lot Number17G1302A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/14/2018
Initial Date FDA Received05/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-