Model Number ESBF2814C103E |
Device Problem
Difficult To Position (1467)
|
Patient Problem
Perforation of Vessels (2135)
|
Event Date 04/11/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
An endurant iis stent graft system was planned to be implanted in a patient for treatment of a 50 mm abdominal aortic aneurysm.However, it was reported that, during the procedure the right external iliac was transected or dissected with the introduction and removal of the device three times.The physician was able to gain proximal control with vessel loops and a 16fr sheaths and completed the evar with a 28mm aui and limb, followed by bilateral femoral endarterectomies and a fem-fem bypass.The physician attributed the cause of the event to diseased external iliac arteries, along with a calcified distal aorta that inhibited the device from passing through the anatomy.No additional clinical sequelae were reported and the patient is fine.
|
|
Manufacturer Narrative
|
Device evaluation summary: device decontaminated with cidex-opa pending further device testing to support the final product analysis findings.The distal tip was curved.Multiple kinks were observed along the stent graft cover.There was no further damage observed to the device.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|