• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL, INC 3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL, INC 3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/11/2018
Event Type  Injury  
Manufacturer Narrative
Vyaire medical reference number (b)(4).At this time, vyaire medical has not received the suspect device from the customer for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
 
Event Description
The customer reported that the end-users heard a noise coming from the ventilator and the ventilator stopped, while in use on a patient.The patient was placed on another ventilator and there was no patient harm associated with this issue.The customer evaluated the ventilator after the incident and confirmed that the ventilator pressurizes, but does not start.
 
Manufacturer Narrative
Results of investigation: vyaire medical performed numerous tests on the suspect component and was not able to duplicate or verify the customer's reported issue during testing and evaluation.The suspect device met all manufacturing specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
kristin graf
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key7503189
MDR Text Key107877036
Report Number2021710-2018-07844
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/11/2018
Initial Date FDA Received05/10/2018
Supplement Dates Manufacturer Received11/20/2018
Supplement Dates FDA Received03/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-