Model Number 3100A |
Device Problems
Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913); Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/11/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Vyaire medical reference number (b)(4).At this time, vyaire medical has not received the suspect device from the customer for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
|
|
Event Description
|
The customer reported that the end-users heard a noise coming from the ventilator and the ventilator stopped, while in use on a patient.The patient was placed on another ventilator and there was no patient harm associated with this issue.The customer evaluated the ventilator after the incident and confirmed that the ventilator pressurizes, but does not start.
|
|
Manufacturer Narrative
|
Results of investigation: vyaire medical performed numerous tests on the suspect component and was not able to duplicate or verify the customer's reported issue during testing and evaluation.The suspect device met all manufacturing specifications.
|
|