• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLIED HEALTHCARE PRODUCTS, INC. GOMCO CIRCUMCISION CLAMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLIED HEALTHCARE PRODUCTS, INC. GOMCO CIRCUMCISION CLAMP Back to Search Results
Model Number 02-02-0501
Device Problem Packaging Problem (3007)
Patient Problem Injury (2348)
Event Date 03/01/2018
Event Type  Injury  
Manufacturer Narrative
Clamps have not been returned for evaluation.We were told the sizes were matched and the clamps were put back into service.Date code on clamps is unknown.If you try to assemble 1.45 bell (stud) with a 1.45 base plate, it cannot be assembled because the bell (stud) falls through the base plate.If you assemble a 1.45 bell (stud) into a 1.3 base plate, the arm will sit at an obvious improper angle and the bell (stud) will hang too far below the base plate for proper use.The size of the clamps is metal stamped on both the base plate and the bell (stud).Parts from different clamps should not be mixed during cleaning and sterilization.We cannot determine if the parts in question were made by gomco or from some other manufacturer from the pictures that were sent in.The instruction manual states the following: warning: this clamp must never be used if component parts are, damaged, missing, clearly worn, or the assembled device does not perform as described.Prior to use, always insure that plate and bell are the same size.Prior to initiating surgical procedure, you must insure that the expected clamping function has been properly achieved.Use only component parts manufactured by "gomco" when assembling this device.Important: some bleeding may occur and/or some sutures may be required depending on the prescribed surgical technique.A chart of bell sizes for comparing bell to size is on page 6 of the manual.
 
Event Description
It was reported that a neonate underwent circumcision by a provider using a gomco clamp.The clamp that was retrieved from sterile packaging had a base and a bell that were different sizes.Despite being different sizes, the instrument was able to be assembled as if the pieces match.In addition, the etching on the instrument indicating the size is on the underside of the equipment, is extremely small, and is the same color as the instrument (silver metal).The neonate sustained a skin injury that required urology consult and revision surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GOMCO CIRCUMCISION CLAMP
Type of Device
CIRCUMCISION CLAMP
Manufacturer (Section D)
ALLIED HEALTHCARE PRODUCTS, INC.
1720 sublette ave
st. louis MO 63110
Manufacturer Contact
jon stillman
1720 sublette ave.
st. louis, MO 63110
3142681616
MDR Report Key7507041
MDR Text Key108005946
Report Number1924066-2018-00002
Device Sequence Number1
Product Code HFX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PREAMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number02-02-0501
Device Catalogue Number02-02-0501
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/10/2018
Initial Date FDA Received05/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age1 DA
Patient Weight3
-
-