Model Number 2090 |
Device Problems
Computer Software Problem (1112); Device Emits Odor (1425); Overheating of Device (1437)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/30/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: analysis confirmed the reported issue; the power supply was electrically out of specification.No other anomalies were found.The power supply was replaced, the hard drive was reconfigured and software reloaded.The programmer passed all final functional and systems tests.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that when the programmer was turned on, it displayed an error.The programmer was turned off and back on, and then began to smoke out of the back.It was noted that the smoke smelled like burning dust.The programmer has been returned for service.There was no patient involvement.
|
|