MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problems Computer Software Problem (1112); Device Emits Odor (1425); Overheating of Device (1437)
Patient Problem No Patient Involvement (2645)
Event Date 04/30/2018
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis confirmed the reported issue; the power supply was electrically out of specification.No other anomalies were found.The power supply was replaced, the hard drive was reconfigured and software reloaded.The programmer passed all final functional and systems tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that when the programmer was turned on, it displayed an error.The programmer was turned off and back on, and then began to smoke out of the back.It was noted that the smoke smelled like burning dust.The programmer has been returned for service.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7507281
MDR Text Key108090925
Report Number2182208-2018-00896
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00613994581570
UDI-Public00613994581570
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/30/2018
Initial Date FDA Received05/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-