• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Catalog Number BI70002000
Device Problems Break (1069); Difficult to Open or Close (2921)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/16/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The representative installed new screws and cable guide.The imaging system then passed the system checkout and was found to be fully functional.No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding an imaging device being used for a spinal fusion procedure.It was reported that the site was attempting to remove the imaging system from around the patient and the door would not open.They attempted to manually open the door, but the cable guide was broken in the process.There was no reported impact on patient outcome.There was a reported delay to the procedure of 10 minutes due to this issue.
 
Manufacturer Narrative
Correction: event date has been updated with new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
stephanie riley
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7509135
MDR Text Key108142089
Report Number1723170-2018-02015
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00643169639683
UDI-Public00643169639683
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/17/2018
Initial Date FDA Received05/11/2018
Supplement Dates Manufacturer Received07/10/2018
Supplement Dates FDA Received07/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/23/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age71 YR
Patient Weight88
-
-