Catalog Number BI70002000 |
Device Problems
Break (1069); Difficult to Open or Close (2921)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
A medtronic representative went to the site to test the equipment.The representative installed new screws and cable guide.The imaging system then passed the system checkout and was found to be fully functional.No parts have been received by the manufacturer for evaluation.
|
|
Event Description
|
Medtronic received information regarding an imaging device being used for a spinal fusion procedure.It was reported that the site was attempting to remove the imaging system from around the patient and the door would not open.They attempted to manually open the door, but the cable guide was broken in the process.There was no reported impact on patient outcome.There was a reported delay to the procedure of 10 minutes due to this issue.
|
|
Manufacturer Narrative
|
Correction: event date has been updated with new information.
|
|