The customer stated that they received erroneous results for three patient samples tested for the elecsys anti-ccp immunoassay (anti-ccp) on a cobas e 411 immunoassay analyzer (e411) and a cobas 6000 e 601 module (e601) used at a second site.It was asked, but it is not known if any erroneous results were reported outside of the laboratory.The first sample initially resulted with an anti-ccp value of 34.7 u/ml (reference range = < 17 u/ml) on the e411 analyzer.The sample was repeated in another laboratory using an unknown test method, resulting as 0.4 u/ml (reference range = 0 - 5 u/ml).The second sample initially resulted with an anti-ccp value of > 500 u/ml (reference range = < 17 u/ml) on the e411 analyzer.The sample was repeated on the e601 analyzer at the second site, resulting as 495 u/ml (reference range = < 17 u/ml).The sample was repeated in another laboratory using an unknown test method, resulting as 240 u/ml (reference range = 0 - 5 u/ml).The third sample initially resulted with an anti-ccp value of 41.7 u/ml (reference range = < 17 u/ml) on the e411 analyzer.The sample was repeated on the e601 analyzer at the second site, resulting as 44 u/ml (reference range = < 17 u/ml).The sample was repeated in another laboratory using an unknown test method, resulting as 1.2 u/ml (reference range = 0 - 5 u/ml).No adverse events were alleged to have occurred with the patients.The e411 analyzer serial number is (b)(4).The serial number of the e601 analyzer was asked for, but not provided.
|
Investigations determined the calibration signals to be ok, but on the low side.Quality control values were normally within specifications, with some outliers.Based on calibration and control data, a general reagent issue can most likely be excluded.According to product labeling, different results can be generated when comparing anti-ccp results from different methodologies.Anti-ccp values determined on patient samples by different testing procedures cannot be directly compared with one another and could be the cause of erroneous medical interpretations.Therefore, the results reported by the laboratory to the physician should include: "the following results were obtained with the elecsys anti-ccp assay.Results from assays of other manufacturers cannot be used interchangeably.".
|