• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM Back to Search Results
Catalog Number 12673-03
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/08/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device was not returned for analysis.The reported difficulties and subsequent treatment appear to be related to an interaction of the device with patient anatomy due to circumstances of the procedure.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling of the device.The other perclose proglide device is filed under a separate medwatch manufacturer report reference number.
 
Event Description
It was reported that an arteriotomy closure of a right common femoral artery was attempted using two proglide devices with a 6f sheath after a cerebral angiogram procedure.Reportedly, a suture break occurred with both proglide devices.An angio-seal was used to achieve hemostasis.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Type of Device
SUTURE MEDIATED CLOSURE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7513031
MDR Text Key108225456
Report Number2024168-2018-03575
Device Sequence Number1
Product Code MGB
UDI-Device Identifier08717648013089
UDI-Public08717648013089
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Catalogue Number12673-03
Device Lot Number8022041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/08/2018
Initial Date FDA Received05/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age87 YR
-
-