Pc-(b)(4).Date sent to the fda: (b)(4) 2018.It was reported performance pull off suture needle.A detached needle was received for evaluation.During the visual inspection of a needle, the swage and attachment area were as to be expected.The hole of the needle was examined under 20x magnification for suture remnant and remnant was found into to hole and marks on body and edge needle that appears to be by use of the surgical instruments.This condition caused the performance - pull off suture needle.The suture was not returned for evaluation to avoid this kind of damage the packages, grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the attachment end to the point.Per the needle condition the assignable cause of the performance - pull off suture needle, was an improper handling.
|