A review of the manufacturing records for the device verified that the lot met all pre-release specifications.The device remains implanted so no engineering investigation could be performed.The image review stated that from the limited fluoro imaging provided the gore helex septal occluders left and right atrial discs appear not to be on the appropriate sides of the atrial septum.A portion of the right atrial disc appears to be on the left atrial side, however, without echocardiography and the limited fluoroscopy imaging provided this cannot be confirmed.A catheter appears to cross through the device from the right atrial side to the left atrial side.This is consistent with a communication between the two atriums still being present.Per the performing facility a cribriform device was implanted and the patient was doing well following the procedure.(b)(4).Udi: (b)(4).
|