Model Number 37602
Device Problems Low Battery (2584); Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problems Therapeutic Effects, Unexpected (2099); Malaise (2359)
Event Date 05/11/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient with an implanted neurostimulator (ins) for parkinsons dual and movement disorders.The caller reported that the patient encountered an unknown error code on the patient programmer and the patient was not feeling well.The caller reported that this was first observed on (b)(6) 2018.The caller called back on (b)(6) 2018 reporting that they checked the patient's ins and the unknown error code was an elective replacement indicator (eri) message.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received from the healthcare provider (hcp) reported the cause of the patient suddenly not feeling well was the battery showing the call the doctor signal, but not having the phone number to call the hcp¿s office.An appointment was scheduled for the patient with a pending replacement.It was noted the patient¿s weight wasn¿t documented as they hadn¿t been set yet.No further complications were anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7522881
MDR Text Key108561809
Report Number3004209178-2018-11174
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00613994761057
UDI-Public00613994761057
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2014
Device Model Number37602
Device Catalogue Number37602
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/14/2018
Initial Date FDA Received05/17/2018
Supplement Dates Manufacturer Received05/17/2018
Supplement Dates FDA Received05/30/2018
Date Device Manufactured05/15/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
-
-