• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROMODULATION ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROMODULATION ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37612
Device Problems Charging Problem (2892); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/26/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer via a manufacturing representative regarding a patient with deep brain stimulation (dbs) im planted for parkinson¿s disease (pd).It was reported that the recharger was not holding it¿s charge.Troubleshooting consisted of trying different plugs; however, it was still not charging the recharger.It was noted that this could be an issue with the wire.As a result, the implantable neurostimulator (ins) was not working for two days.The issue was not resolved.There were no patient symptoms or surgical intervention reported or anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from a manufacturing representative.It was reported that it was not confirmed if the patient's ins had depleted.The patient tried to charge for two days but they were unable to.The replacement recharger resolved the issue.
 
Manufacturer Narrative
Other applicable components are : product id: 37651, serial#: (b)(4), product type: recharger.Analysis of the patient's recharging system found the connector pin was bent.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7526129
MDR Text Key108671818
Report Number3007566237-2018-01495
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/26/2018
Initial Date FDA Received05/18/2018
Supplement Dates Manufacturer Received05/18/2018
08/20/2018
Supplement Dates FDA Received06/07/2018
09/12/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-