Model Number 801188 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/25/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the manufacturer's subsidiary, the issue was discovered while testing the unit in the service center at the user facility.The reported issue was lost after 30 seconds and they were unable to duplicate it.
|
|
Event Description
|
It was reported that during the use of the device for a non-clinical activity, the electronic patient gas system (epgs) was displaying as an x on the central control monitor.There was no patient involvement.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed the electronic patient gas system (epgs) to cause a red x on the central control monitor (ccm) was because of a defective generic board.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|