|
Model Number 8098027 |
Device Problems
Display or Visual Feedback Problem (1184); Device Issue (2379)
|
Patient Problem
Radiation Exposure, Unintended (3164)
|
Event Date 04/18/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Siemens conducted a thorough evaluation of the reported event.The investigation concluded that by performing a new checkup of the system, a gain-calibration was done which optimizes the function of the modules in the detector.The dicom images that were provided show that the mentioned artifacts are clearly visible so that the risk of a misdiagnosis can be excluded.According to all of the information provided, this solved the issue.The customer/user of the system is reminded to perform daily check ups of the system as recommended in the user manual.No systematic product issue was detected, and no further measures were required.
|
|
Event Description
|
Siemens was made aware on (b)(6) 2018, that during a patient procedure using the somatom definition as, a (b)(6) year old male patient got rescanned due to artifacts appearing in the ct scan.The artifacts are described as a dark rotating shadow.The patient was rescanned with additional x-ray dose.There was no negative impact reported, no further injury to the patient.This event occurred in (b)(6).
|
|
Search Alerts/Recalls
|
|
|