• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that the system froze during a procedure and the footswitch cable had a blister.Another system was used to complete the procedure.There was no patient impact reported.
 
Manufacturer Narrative
The system was examined and the reported event was not replicated.However, the footswitch cable (which belongs to the system) was replaced (and returned for evaluation), as a preventive measure.The system was manufactured on june 24, 2014.Based on qa assessment, the product met specifications at the time of release.A footswitch cable was received for evaluation.A visual assessment of the returned sample revealed a bump in the cable near the footswitch-end connector.The sample was installed onto a calibrated system for functional testing.The sample was found to meet relevant specifications.The system was found to meet specifications.Therefore, the root cause of the reported freezing issue cannot be determined conclusively.Although the cable was found to be visually nonconforming, the sample was found to meet relevant specifications when testing.Therefore, the root cause of the footswitch cable issue cannot be determined conclusively.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7532153
MDR Text Key108992655
Report Number2028159-2018-01066
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Other Device ID Number380657511501
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/24/2018
Initial Date FDA Received05/22/2018
Supplement Dates Manufacturer Received08/02/2018
Supplement Dates FDA Received08/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-