VYAIRE MEDICAL, INC 3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/25/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).At this time, vyaire medical has not received the suspect component from the customer for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
 
Event Description
The customer reported that the ventilator stopped running while in use on a patient.There was no patient harm/injury associated with this issue.
 
Manufacturer Narrative
Results of investigation: a visual inspection of the module did not show any signs of physical damage.After installing the driver assembly into a known good top level unit and after a successful circuit pressure check-out, the unit cycled.After cycling the unit a minimum of 12 hours, vyaire was unable to reproduce the reported issue of overheating or the reported issue of the unit not cycling.All test points read the proper voltages with the driver running.This unit met all specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
avery foster
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key7535097
MDR Text Key108949749
Report Number2021710-2018-07889
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/25/2018
Initial Date FDA Received05/22/2018
Supplement Dates Manufacturer Received08/30/2018
Supplement Dates FDA Received09/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-