• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA SJM TRIFECTA VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA SJM TRIFECTA VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number TF-21A
Device Problem Material Perforation (2205)
Patient Problem Aortic Insufficiency (1715)
Event Type  Injury  
Manufacturer Narrative
An event of aortic insufficiency was reported.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
On an unknown date, a 21 mm trifecta valve was implanted.Seven months post-implant, the patient developed moderate aortic insufficiency.Over the next several months, aortic insufficiency progressed to severe.The echocardiographic studies demonstrated that the leak was not para-valvular, but related to the prosthesis itself.The valve was explanted and replaced with an unknown valve.The leaflet perforation was reported to be secondary to a medially directed cor-knot.The affected patient had a small annulus, and narrow outflow track, which may have contributed to the cor-knot perforating the leaflet when it was in the open position during systole.The status of the patient is unknown.No further information is expected to be received.This information was obtained from a published case study.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM TRIFECTA VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7536658
MDR Text Key108999415
Report Number3008452825-2018-00158
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,litera
Reporter Occupation Physician
Type of Report Initial
Report Date 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberTF-21A
Device Catalogue NumberTF-21A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/27/2018
Initial Date FDA Received05/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-