DEPUY ORTHOPAEDICS INC US ALTRX NEUT 36IDX56OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Model Number 1221-36-056
Device Problems Disassembly (1168); Device Dislodged or Dislocated (2923); Noise, Audible (3273)
Patient Problems Pain (1994); Joint Dislocation (2374); No Known Impact Or Consequence To Patient (2692); Not Applicable (3189); No Code Available (3191); Swelling/ Edema (4577)
Event Date 04/25/2018
Event Type  Injury  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Der states that the poly disassociated from cup.  doi: (b)(6) 2016; dor: (b)(6) 2018; right hip.
 
Manufacturer Narrative
(b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
Litigation records received alleging that one year and four months after the primary surgery, the patient's right hip popped after bending.Diagnostic testing (according to the litigation record) has indicated that the humeral head was riding high in the acetabular compartment and that there was a dislocation.The patient was forced to undergo a head and liner exchange.The records further allege severe and permanent injury, pain, suffering, disability and the need for a revision surgery.Doi: (b)(6) 2016 - dor: (b)(6) 2018; (right hip).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy as not synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
After review of the medical records the patient was revised to address failed right tha with liner dissociation.Pop sensation of the hip, grinding and squeaking-type sensation and pain after bending over.However, xray reported no evidence of hip dislocation.Ct scan confirmed evidence of liner dissociation.Operative note reported capsule under tension with fluid, this was black tarnished from the ceramic on metal contact.There was thick blackened synovium in the joint that were debrided.There was difficulty implanted the head due to native retroversio.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: e3 initial reporter occupation: lawyer.Added: a4, b5, b7, d10 and h6 (patient).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: a2 (age), e1, e2, e3 (removed), g1, g2 and h6 (patient, device).Replaced surgical intervention with device revision or replacement.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRX NEUT 36IDX56OD
Type of Device
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7538184
MDR Text Key109052862
Report Number1818910-2018-60371
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295016342
UDI-Public10603295016342
Combination Product (y/n)N
PMA/PMN Number
K062148
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Model Number1221-36-056
Device Catalogue Number122136056
Device Lot NumberH00457
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/25/2018
Initial Date FDA Received05/23/2018
Supplement Dates Manufacturer Received06/20/2018
11/18/2019
02/17/2020
06/17/2021
08/05/2021
Supplement Dates FDA Received06/27/2018
12/12/2019
02/20/2020
07/05/2021
08/10/2021
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
Treatment
APEX HOLE ELIM POSITIVE STOP; DELTA CER HEAD 12/14 36MM +12; PINN CAN BONE SCREW 6.5MMX25MM; PINN SECTOR W/GRIPTION 56MM; TRI-LOCK BPS SZ 3 STD OFFSET
Patient Outcome(s) Required Intervention;
Patient Age59 YR
Patient Weight93
-
-