MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICM5_13.2
Device Problems Positioning Failure (1158); No Apparent Adverse Event (3189)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/06/2018
Event Type  malfunction  
Manufacturer Narrative
This product is not marketed in the us.Product evaluation found that the lens was returned dry in a micro-centrifuge vial.Visual inspection found the haptic torn, residue on lens, and torn piece missing.No similar complaint type events were reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 13.2 mm vicm5_13.2, -9.50 diopter, implantable collamer lens in the patient's right eye (od) on (b)(6) 2018.During implantation, the foam tip caught the optic and haptic and; causing a tear to the lens.The lens was intraoperatively exchanged with another lens and the problem was resolved.The patient is okay.
 
Manufacturer Narrative
Corrected data: correct patient identifier is (b)(6).(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key7539113
MDR Text Key109109544
Report Number2023826-2018-00761
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date01/31/2020
Device Model NumberVICM5_13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/24/2018
Initial Date FDA Received05/23/2018
Supplement Dates Manufacturer Received07/27/2018
Supplement Dates FDA Received07/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age22 YR
-
-