Model Number M8007A
Device Problem Device Alarm System (1012)
Patient Problem Cyanosis (1798)
Event Date 05/02/2018
Event Type  Injury  
Manufacturer Narrative
There was no product malfunction identified, the device was working as intended.The alarm behavior was accurate and the device worked as designed and configured.The customer was instructed accordingly.No further investigation or action is warranted.
 
Event Description
The customer reported delayed alarms on mp70 on (b)(6) 2018 at approx.11 am.Spo2 and heart rate are the vitals that the alarms are having a delay in sounding.The patient turned blue, was bagged and respiratory treatment was administered, afterwards the patient was stable again.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MP70 INTELLIVUE PATIENT MONITOR
Type of Device
PATIENT MONITOR
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
hewlett-packard str.2
boeblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
betty harris
hewlett-packard str.2
boeblingen 71034
GM   71034
MDR Report Key7540651
MDR Text Key109135980
Report Number9610816-2018-00130
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K021300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM8007A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/02/2018
Initial Date FDA Received05/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age7 MO
Patient Weight6
-
-