• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ SALINE SYRINGE; SALINE FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ SALINE SYRINGE; SALINE FLUSH Back to Search Results
Catalog Number 306574
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/04/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Investigation results: summary: a total of 15 photos were received for evaluation by our quality engineers.Photos show the case label and the shelf box label confirming the lot#7291824.Photos show a shelf box which experienced excessive compression force.There are other photos showing syringes in their packaging flow wrap.The packaging flow wrap is torn on the plunger rod thumb press side.Failure mode is verified.Based on the photos the packaging was likely exposed to excessive compression force inducing damage to the shelf box and packaging flow wrap during handling and transportation.A device history review was completed and there was no documentation of issues for the complaint of batch (b)(4) during this production run.There were no quality notifications issued during the production of this batch listed in the complaint.Conclusion: investigation comments: all our inspections and testing performed while manufacturing this batch were accepted.No rejections were documented.(b)(6) 2018; 15 photos were received.Photos show the case label, the shelf box label confirming the lot#7291824.Photos show a shelf box which experienced excessive compression force.There are other photos showing syringes in their packaging flow wrap.The packaging flow wrap is torn on the plunger rod thumb press side.Root cause: root cause could not be determined.There were no qns issued during the production of this batch listed in the complaint.All inspections and testing were accepted during the production of this batch.Possible root cause.Handling/ transportation.Based on the photos the packaging was exposed to excessive compression force inducing damaged to the shelf box and packaging flow wrap.
 
Event Description
It was reported some of individual packaging was found open on a box of 5 ml bd posiflush¿ saline syringes prior to use.There was no report of exposure, serious injury or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ SALINE SYRINGE
Type of Device
SALINE FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7541444
MDR Text Key109304552
Report Number1911916-2018-00231
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
K011982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,other
Reporter Occupation Other
Type of Report Initial
Report Date 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/31/2020
Device Catalogue Number306574
Device Lot Number7291824
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/04/2018
Initial Date FDA Received05/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-