Model Number FUSION COMPACT |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/30/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved in this concern.Unique device identification (udi) is unavailable.The portable axiem system was returned to the manufacturer for evaluation.It was reported the alleged issue was confirmed.The axiem unit was connected to a test system with the ent application.Test system reported the axiem box and emitter were showing a communication error.The eminter face reported a fault on coils 1, 2, and 3.Device manufacturing date is unavailable.
|
|
Event Description
|
Medtronic received information regarding a navigation system.It was reported that when a manufacturer representative plugged in the emitter, there was an emitter and axiem communication error.It was reported that when the emitter was disconnected, the axiem had green communication.The representative rebooted the system while the emitter was disconnected from the axiem box, but this did not resolve the issue.This issue was discovered outside of a procedure and there was no patient present when this issue was observed.
|
|
Manufacturer Narrative
|
Correction: product and related fields updated to proper value.
|
|