Model Number S7
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Erratic or Intermittent Display (1182)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/01/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.Udi not available for this system at time of filing.Device manufacturing date is unavailable.The software investigation found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.
 
Event Description
Medtronic received information regarding a navigation device being used for a functional endoscopic sinus surgery (fess) procedure.The event occurred intraoperatively during the navigate task and caused minimal delay to surgical time.It was reported that while navigating, the surgeon verified the straight probe and a blue screen appeared.The manufacturer representative (mr) restarted the system and the issue was resolved.The mr also re-verified the probe and the issue was resolved.Navigation and imaging were not aborted.The surgery was completed and there was no impact on patient outcome.
 
Manufacturer Narrative
Device manufacturing date was updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
sarah olhoft
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7544904
MDR Text Key109280597
Report Number1723170-2018-02289
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/01/2018
Initial Date FDA Received05/25/2018
Supplement Dates Manufacturer Received06/06/2018
Supplement Dates FDA Received06/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/20/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse