ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A scrub technician reported that during a surgical procedure the purge cycle on the auto gas fill (agf) cycle did not work.The staff was able to mix c3f8 by manual process.There was no harm to patient reported.
 
Manufacturer Narrative
Additional information is provided.The company service representative examined the system but was unable to replicate any issue related to the reported event.The pneumatics module was replaced as a preventative measure.The system was then tested and met all product specifications.The system was manufactured on march 5, 2010.Based on qa assessment, the product met specifications at the time of release.The system was found to meet specifications; therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7545948
MDR Text Key109493753
Report Number2028159-2018-01096
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/04/2020
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Other Device ID Number380657511501
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/03/2018
Initial Date FDA Received05/25/2018
Supplement Dates Manufacturer Received07/27/2018
Supplement Dates FDA Received08/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/05/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-