MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION Back to Search Results
Model Number RA002-4545SL70
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Information (3190)
Event Type  Injury  
Event Description
It was alleged that one of the patient's magec rod would not distract.The physician revised the rods without incident.
 
Manufacturer Narrative
A visual inspection of the returned rods revealed that the rods were partially distracted with score marks on the distraction rods.The rods could be lengthened by hand when pulling on the distraction rod, which confirms the pin failures.A review of the lot history record for the device revealed that the device met all of the required quality inspections and that the products were released within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise, suite 100
aliso viejo CA 92656
MDR Report Key7547170
MDR Text Key109360845
Report Number3006179046-2018-00041
Device Sequence Number1
Product Code PGN
Combination Product (y/n)N
PMA/PMN Number
K140613
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberRA002-4545SL70
Device Catalogue NumberPA0477
Device Lot NumberA140804-01-00
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/26/2018
Initial Date FDA Received05/26/2018
Supplement Dates Manufacturer Received04/26/2018
Supplement Dates FDA Received09/18/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-