Model Number 16402
Device Problem Device Displays Incorrect Message (2591)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 05/09/2018
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) was unable to duplicate the reported complaint.He found the optical sensors on the dual tachometer board were coated with white dust.He replaced the dual tachometer board.The unit operated to the manufacturer specifications.The suspect part was returned to the manufacturer for further evaluation.This complaint is related to (b)(4) / medwatch #1828100-2018-00262.
 
Event Description
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, a "belt slip" error was displayed.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed residue found on the opto-switches likely from the pumps encoder wheel causing the belt slip errors.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
Additional information was received that the surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per clinical review: during a cardiopulmonary bypass (cpb) procedure on (b)(6) 2018, the roller pump gave the perfusionist a belt slip error, along with a thumping noise.The perfusionist opted to turn off the roller pump temporarily and then restart it.This was an ancillary roller pump so she was able to do this without concern.The thumping continued, but the belt slip error resolved itself and she was able to use the roller pump for the remainder of the procedure without a functional issue.The incident did not delay the surgical procedure, and there was no harm or blood loss associated with the occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 UNIVERSAL ROLLER PUMP
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7550487
MDR Text Key109494634
Report Number1828100-2018-00281
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K953901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 08/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/09/2018
Initial Date FDA Received05/29/2018
Supplement Dates Manufacturer Received06/07/2018
07/20/2018
Supplement Dates FDA Received06/28/2018
08/01/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-