• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problems No Display/Image (1183); Loss of Power (1475)
Patient Problem No Patient Involvement (2645)
Event Date 04/17/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a faulty processor board.The technician replaced the board to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The board has been requested for return to livanova (b)(4) for further investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that an s5 roller pump display lost power during maintenance.There was no patient involvement.
 
Manufacturer Narrative
During the investigation of the processor board the reported issue could be reproduced.Evaluation of the serial read out of the pump memory showed internal firmware bugs in the microcontroller.As part of the devices fw upgrade process, the memory must be deleted prior to installing the new fw version.In the unit involved in this event, this step of clearing the memory was not properly followed causing the presence of bugs on the failing board.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Event Description
See initial.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key7553453
MDR Text Key109750349
Report Number9611109-2018-01003
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/25/2018
Initial Date FDA Received05/30/2018
Supplement Dates Manufacturer Received05/10/2019
Supplement Dates FDA Received06/07/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-