STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO12.6
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Edema (1820); No Code Available (3191)
Event Date 04/04/2018
Event Type  Injury  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.Device evaluation: the lens was returned dry, in micro-centrifuge vial.Visual inspection found no visible damage to the lens and there was residue on the lens.(b)(4).
 
Event Description
The reporter indicated the surgeon implanted a 12.6mm vticmo12.6 implantable collamer lens, -14.00/+3.0/094 (sphere/cylinder/axis), in the patients left eye (os), on (b)(6) 2018.The lens was explanted on (b)(6) 2018 due to patient related factor of vertigo.The vertigo was pre-existing to icl implantation.At one day post-op, the crystalline lens was clear and there was 1+ incisional edema.The vertigo has been reduced and patient will return to c/l wear.The patient elected to have the icl explanted.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key7555304
MDR Text Key109612290
Report Number2023826-2018-00832
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date08/31/2020
Device Model NumberVTICMO12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/08/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/30/2018
Initial Date FDA Received05/30/2018
Supplement Dates Manufacturer Received08/01/2018
Supplement Dates FDA Received08/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age23 YR
-
-