Model Number S7
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Patient Involvement (2645)
Event Date 05/02/2018
Event Type  malfunction  
Manufacturer Narrative
Udi not available for this system at time of filing.Device manufacturing date is unavailable.No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding a navigation system.It was reported that during case setup, after importing images, the system became unresponsive.The touchscreen did not respond to user input.A manufacturer representative was forced to hard shutdown the system and restart.The issue was resolved following the restart.This issue was discovered outside of a procedure and there was no patient present when this issue was observed.
 
Manufacturer Narrative
Additional information: unique device identification (udi) and device manufacture date provided.Correction: product and related fields updated to proper value.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7557203
MDR Text Key109742739
Report Number1723170-2018-02328
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 07/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9735665
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/02/2018
Initial Date FDA Received05/31/2018
Supplement Dates Manufacturer Received06/11/2018
Supplement Dates FDA Received07/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-