U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION Back to Search Results
Model Number MS1-4590S
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/02/2018
Event Type  Injury  
Event Description
It was alleged that the patient's magec rods would not distract.The physician revised the rods with new magec rods without incident.
 
Manufacturer Narrative
Rod 1 lot # 150128-10: a visual inspection of the returned rod revealed it was partially distracted.Functional testing revealed that the rod could distract and retract with the manual distractor.It was determined the rod was functional.Rod 2 lot # 150428-13: a visual inspection of the returned rod revealed it was partially distracted.Functional testing revealed the rod could not be distracted or retracted using the manual distractor.The rod was able to be distracted and retracted with the erc.The rod was observed to have score marks and debris build up, which may have caused the reduced functionality of the rod.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise, suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise, suite 100
aliso viejo CA 92656
Manufacturer Contact
carol bleakley
101 enterprise, suite 100
aliso viejo, CA 92656
MDR Report Key7562944
MDR Text Key109892588
Report Number3006179046-2018-00043
Device Sequence Number1
Product Code PGN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140613
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberMS1-4590S
Device Catalogue NumberPA0516
Device Lot NumberA141118-01, A141210-01
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/02/2018
Initial Date FDA Received06/01/2018
Supplement Dates Manufacturer Received05/02/2018
Supplement Dates FDA Received09/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age9 YR
-
-