• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. PICO WITH SOFT PORT 15X15CM; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. PICO WITH SOFT PORT 15X15CM; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66801361
Device Problem Pumping Problem (3016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/01/2018
Event Type  malfunction  
Event Description
It was reported that on the third day of the npwt utilizing a pico, the pump stopped working.The treatment was continued with a back-up device.No patient harm reported.
 
Manufacturer Narrative
Our investigation into this complaint has now concluded.The returned pump was passed to our in house experts so a thorough technical analysis could be carried out to help to identify if there were any problems with the pump and/or what could have caused the reason for the complaint.A review of the associated batch manufacturing records did not identify any issue at the point of manufacture which may have caused or contributed to the issue highlighted by the complainant.In addition it can be confirmed that all finished product specification testing was satisfied at the point of release.Our experts have provided a report of the analysis undertaken which confirms that ¿the device was tested and no anomalies was found¿ the pump had reached its 7 day expiry with a system time of 7 days, 8hrs.The eol (end of life) flag was reset and the pump functioned as expected with no issues detected.Prior to manufacturing, all pico pumps undergo full functionality testing to an agreed specification.This ensures that all pumps are working correctly before being shipped to our customers; any failures identified are segregated for investigation.The 7 day timer is activated as soon as the batteries are inserted.As stated in the ifu ¿after 7 days of therapy the pump will automatically cease functioning¿ smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.We will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PICO WITH SOFT PORT 15X15CM
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 8N
UK  HU3 28N
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 8N
UK   HU3 28N
Manufacturer Contact
sarah poettker
101 hessle road
hull HU3 2-BN
UK   HU3 2BN
MDR Report Key7567614
MDR Text Key110033741
Report Number8043484-2018-00126
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K112127
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66801361
Device Lot Number1721
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/07/2018
Initial Date FDA Received06/04/2018
Supplement Dates Manufacturer Received05/07/2018
Supplement Dates FDA Received06/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-