Model Number S7
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/09/2018
Event Type  malfunction  
Manufacturer Narrative
Provided to test new demographic fields in production and not related to actual data.Udi number unavailable.Device manufacturing date unavailable.Software logs received and currently under analysis.
 
Event Description
Medtronic received information that, while in an electrode placement deep brain stimulation (dbs) procedure, the navigation system display reverted to the medtronic logo page and restarted without prompt from the user.There was a reported delay to the procedure of less than 1 hour due to this issue.There was no reported impact on patient outcome.No additional information was provided.
 
Manufacturer Narrative
The software logs were reviewed but provided no additional insight regarding the cause of the reported complaint.The software investigation found that a probable cause was unable to be determined without further information since the on-going investigation proved to be inconclusive based on the information provided.
 
Manufacturer Narrative
Correction: suspect device information provided.Additional information: device manufacturing date provided.Additional information: a medtronic representative went to the site to test the equipment.The imaging system then passed the system checkout and was found to be fully functional.The software investigation found that the reported event was consistent with an existing software investigation.This issue was documented in a medtronic navigation software anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
amos jarrette
navigation customer quality
826 coal creek circle
louisville, CO 80027
7208903200
MDR Report Key7568706
MDR Text Key110050996
Report Number1723170-2018-02512
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733927
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/09/2018
Initial Date FDA Received06/05/2018
Supplement Dates Manufacturer Received06/07/2018
06/21/2018
Supplement Dates FDA Received07/05/2018
07/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age17 YR
Patient Weight68
-
-