Model Number 8637-40
Device Problems Occlusion Within Device (1423); Aspiration Issue (2883)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/04/2018
Event Type  Injury  
Manufacturer Narrative
The main component of the system.Other relevant device(s) are: product id: 8780, serial/lot # :(b)(4) ubd: (b)(6) 2016, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider (hcp) via a device manufacturer representative regarding a patient receiving baclofen (100 mcg/ml, 39.97 mcg/ml), dilaudid (15.0 mg/ml, 5.995 mg/day) and bupivacaine (7.5 mg/ml, 2.9975 mg/day) via an implantable infusion pump for the treatment of non-malignant pain and chronic low back pain.It was reported that the hcp was unable to aspirate catheter due to clogged catheter.They tried to trim catheter to clear blockage and the catheter was replaced with a new 8780 catheter.There was no environmental/external/patient factors that may have led or contributed to the issue.The issue was resolved and patient status was alive with no injury at the time of the report.No further complication was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7568765
MDR Text Key110052921
Report Number3004209178-2018-12566
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/28/2012
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/04/2018
Initial Date FDA Received06/05/2018
Date Device Manufactured06/03/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age47 YR
-
-