Device updated by mfr.: the device was received in two separate pieces for analysis.One section consisted of hub/manifold and hypotube and the other section of hypotube, shaft and balloon.A visual and tactile examination identified a complete break approximately 173mm distal to the strain relief on the hypotube of the device.Multiple severe kinks were also noted along the length of the hypotube.This type of damage is consistent with excessive force being applied to the delivery system.No issues were noted with the hypotube that may have contributed to the complaint incident.A visual and microscopic examination identified no damage to any of the blades.All blades were present and fully bonded to the balloon material.No issues were noted with the blades or pads that could have contributed to the complaint incident.Oxidation was noted on the blades.The balloon was unfolded and solidified saline solution was noted within the balloon material which indicates it had been subjected to positive pressure.Due to the condition of the returned device is was not possible to inflate the balloon, however a full visual and microscopic examination identified no damage or any tears within the balloon material.No issues were noted with the balloon material that could have contributed to the complaint incident.A visual and microscopic examination identified no damage or any issues with the tip or markerbands of the device.No other issues were identified during the product analysis.The investigation conclusion was unable to be determined.(b)(4).
|