MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC HD 1000I SHEARS 36CM SHAFT; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC HD 1000I SHEARS 36CM SHAFT; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARHD36
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/12/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # r92513.Investigation summary: the device was returned with the tissue pad damaged, melted and 100% present.The device was connected to a gen11 and the device did activate during functional testing.When device was disassembled to inspect the internal components, no anomalies were found related to the reported event.However, it was found that there was a cracked internal component; the y-link.Probable causes of tissue pad damage are applying pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The batch history record was reviewed and no defects, nc¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that during a laparoscopic assisted vaginal hysterectomy and bilateral salpingo-oophorectomy procedure, the tissue pad came loose.The customer stated that it did not come loose due to overuse or melting.No piece of the pad fell off.Procedure was completed with another device of the same product code.There were no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC HD 1000I SHEARS 36CM SHAFT
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7570086
MDR Text Key110181223
Report Number3005075853-2018-10308
Device Sequence Number1
Product Code LFL
UDI-Device Identifier20705036015052
UDI-Public20705036015052
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160752
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Catalogue NumberHARHD36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/17/2018
Initial Date FDA Received06/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-