MAUDE Adverse Event Report: BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD NEXIVA¿ CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD NEXIVA¿ CLOSED IV CATHETER SYSTEM; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383532
Device Problem Sticking (1597)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/17/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that during use a bd nexiva¿ closed iv catheter system malfunctioned as "the needle become stuck on removal, which then resulted in us having to fully remove the cannula and re-cannulate the child again." there was no report of exposure, injury or medical intervention needed.
 
Event Description
It was reported that during use a bd nexiva¿ closed iv catheter system malfunctioned as "the needle become stuck on removal, which then resulted in us having to fully remove the cannula and re-cannulate the child again." there was no report of exposure, injury or medical intervention needed.
 
Manufacturer Narrative
Device/batch history record review: the lot was built and packaged on nfa line 2 from 20oct2017 through 25oct2017 for the quantity of (b)(4).All required challenge samples and testing during the build of the catheters was performed per specifications, in accordance with the in-process sampling plans.A related qn (see attached) for ¿failed retraction¿ was initiated; disposition of the product, root cause and corrective action were applied according to the quality control plan.The peura (end user risk analysis) was analyzed to determine the risk to customer.The analysis showed that due to low occurrence, current risk is acceptable.Observations and testing: received 2 nexiva 22ga units.Unit 1: received an unused unit where all the components were present.The unit was received along with a piece of top web (packaging) from lot number: 7290663.Unit 2: received a used unit where the needle was fully disengaged but the catheter was still attached to the tip shield.The unit was received along with a piece of top web (packaging) from lot number: 7290663.Simulation test (disengagement): unit 1: simulation test was performed and successfully disengaged and the adapter was released without resistance.No adhesive or physical-mechanical damage was found on any of the components of the unit.Unit 2: simulation test was performed and although the cannula successfully disengaged without resistance and the safety mechanism was applied, the adapted stayed attached to the tip shield.Visual/microscopic examination: revealed traces of cured adhesive on several locations throughout the tip shield preventing the adapter from decoupling.The needle fully disengaged and the safety mechanism functioned as expected.The failure described in the incident report was caused when cured adhesive ¿bonded¿ the tip shield and adapter preventing it from decoupling.Per qn the failure was initiated due to machine wear (o rings worn out).Corrective action was applied.A formal corrective action will not be initiated at this time.Customer complaint trends are evaluated on a monthly basis.If the trend of a specific type of complaint warrants a formal corrective action, resources will be assigned at that time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD NEXIVA¿ CLOSED IV CATHETER SYSTEM
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
MDR Report Key7570540
MDR Text Key110346214
Report Number1710034-2018-00262
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier30382903835325
UDI-Public30382903835325
Combination Product (y/n)N
PMA/PMN Number
K161777
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 08/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date09/30/2020
Device Catalogue Number383532
Device Lot Number7290663
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/31/2018
Initial Date Manufacturer Received 05/17/2018
Initial Date FDA Received06/05/2018
Supplement Dates Manufacturer Received05/17/2018
Supplement Dates FDA Received08/08/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-