TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRUSYSTEM 7000 DV Back to Search Results
Model Number 1723633
Device Problems Device Operates Differently Than Expected (2913); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/10/2018
Event Type  malfunction  
Manufacturer Narrative
A trumpf medical service technician was dispatched to investigate the incident and found that fluid had gotten into the rc port and damaged the motor control board.The motor control board and keyboard were replaced, and the table operated as intended.
 
Event Description
A user facility alleged that a trumpf medical trusystem 7000 dv operating table could not be moved with the remote or column controls while a patient was positioned on the table top.The emergency override would not activate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUSYSTEM 7000 DV
Type of Device
7000 DV
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld, 07318
GM  07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld, 07318
GM   07318
MDR Report Key7574894
MDR Text Key110587641
Report Number3007143268-2018-00007
Device Sequence Number1
Product Code JEA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Remedial Action Repair
Type of Report Initial
Report Date 05/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1723633
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/10/2018
Initial Date FDA Received06/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-