• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) INST KIT PROBE THORACIC ASSY; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) INST KIT PROBE THORACIC ASSY; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9734913
Device Problem Bent (1059)
Patient Problem No Patient Involvement (2645)
Event Date 05/14/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Lot no.And manufacture date not available for this instrument at time of filing.No parts have been returned to the manufacturer for evaluation.
 
Event Description
Medtronic received information that, while outside of a procedure, the thoracic probe for the site was bent.There was no patient present when this issue was identified.No additional information was provided.
 
Manufacturer Narrative
Additional information: device lot number provided.The probe was returned to the manufacturer for evaluation.Testing found that the tip of the probe was not bent.However, it was noted that impact marks at the back end led to fit issues with trackers.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.
 
Manufacturer Narrative
Additional information: unique device identification (udi) and device manufacture date provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INST KIT PROBE THORACIC ASSY
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7575254
MDR Text Key110333547
Report Number1723170-2018-02550
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00643169636095
UDI-Public00643169636095
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K124004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9734913
Device Lot Number170606
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/14/2018
Initial Date FDA Received06/06/2018
Supplement Dates Manufacturer Received08/07/2018
08/28/2018
Supplement Dates FDA Received08/28/2018
09/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-