Catalog Number 05170478001 |
Device Problem
Low Test Results (2458)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/15/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred in (b)(6).Unique identifier (udi)#: (b)(4).
|
|
Event Description
|
The customer complained of an erroneous low potassium (k) result from the cobas b 123 instrument.The initial k result from the b 123 instrument was 4.38 mmol/l.The same sample was tested on a beckman au5800 instrument and the k result was 7.08 mmol/l.The results were reported outside of the laboratory and provided to the patient.There was no allegation of an adverse event.The k electrode lot number was 21580496 with an expiration date of 02-jul-2018.Qc results were acceptable.
|
|
Manufacturer Narrative
|
Based on investigation of data logs from the b 123 instrument, the alleged k result of 4.38 mmol/l was measured on 15-may-2018.The k result from the beckman instrument was measured on 16-may-2018.The difference in k results from the b 123 instrument and the beckman instrument was due to comparing blood samples measured on 2 different days.An analysis of sensor data shows the slope of the sensor was well within limits and no abnormal signals were observed between 05-may-2018 and 18-may-2018.Based on the data provided, the cobas b 123 instrument functioned correctly.A product problem was not found.
|
|
Search Alerts/Recalls
|