• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH ULTRA DUO HIGH FLUID CART NON-STERILE INSTRUCTIONS FOR USE; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH ULTRA DUO HIGH FLUID CART NON-STERILE INSTRUCTIONS FOR USE; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/04/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).K081047; k123188; k133786.The investigation is still in progress.Once the investigation is complete a follow up mdr will be submitted.
 
Event Description
It was reported that the unit was leaking fluid from the bottom.The event occurred during surgery.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Udi: (b)(4).No functional defect was found with the device during evaluation, therefore, the complaint will be processed using icwr0008 rev.1.As such, no device history record and complaint history review are required for the complaint.Using crm to query for serial number (b)(4), the device was noted to have not been previously repaired by zimmer biomet surgical.On 04 june 2018, it was reported from community hospital south that a unit was leaking fluid from the bottom during.Lmc service solutions was contacted about the cart and dispatched a service technician to be at the site.The technician arrived at the site and was unable to confirm the reported event and noted not issue with the cart.The device was tested, inspected, and repaired as per cl ¿ repair carts rev.4.Service work order (b)(4) on 04 june 2018.The reported event was never confirmed during inspection of the device and the device was noted to be functioning as intended.Therefore the root cause could not be determined.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA DUO HIGH FLUID CART NON-STERILE INSTRUCTIONS FOR USE
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
MDR Report Key7579849
MDR Text Key110457476
Report Number0001954182-2018-00036
Device Sequence Number1
Product Code JCX
Combination Product (y/n)N
PMA/PMN Number
PK162421
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 06/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00514010100
Device Lot Number0027358
Other Device ID Number(01)00889024465992
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/04/2018
Initial Date FDA Received06/07/2018
Supplement Dates Manufacturer Received06/04/2018
06/13/2018
Supplement Dates FDA Received06/11/2018
06/14/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-