The reported event was confirmed during functional testing, the autopulse platform did not power on due to a defective power switch cable.Review of the retrieved archive data found no significant discrepancies related to the reported complaint.As part of routine service during testing, the platform was examined and found unrelated to the reported complaint, damaged front enclosure, a bent battery lock, and an lcd screen with no backlight.Front enclosure, a battery lock, and an lcd were replaced to address the issue.The autopulse platform is a reusable device and was manufactured in 2013.This type of physical damage found during visual inspection is characteristic of normal wear and tear for the age of the device.Following the service and repair, the platform was tested for 15 minutes with large resuscitation testing fixture, (lrtf) equivalent to 250 pound patient and passed successfully with no issue or faults observed.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse with serial number (b)(4).
|