BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLAMAP ORION¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number M004RC64S0
Device Problem Device Sensing Problem (2917)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/27/2018
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: visual inspection showed that the returned orion the deployment array is separating.The electrical test was performed and the device failed the test.Magnetic sensor resistance and inductance testing revealed both pairs are out of specification.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion was unable to be determined.(b)(4).
 
Event Description
Reportable upon analysis completed on 11may2018: it was reported that tracking issues occurred.A intellamap orion was selected for a ventricular tachycardia ablation.However after the orion was inserted into the left ventricle the rhythmia system showed an error about the magnetic tracking sensor on the orion catheter.The procedure was completed using another orion catheter.Upon analysis it was noticed that the deployment array was separating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLAMAP ORION¿
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7583378
MDR Text Key110577408
Report Number2134265-2018-04845
Device Sequence Number1
Product Code DRF
UDI-Device Identifier08714729841968
UDI-Public08714729841968
Combination Product (y/n)N
PMA/PMN Number
K122461
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/15/2018
Device Model NumberM004RC64S0
Device Catalogue NumberRC64S
Device Lot Number21392000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/11/2018
Initial Date FDA Received06/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/15/2017
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-