• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750833
Device Problems Device Inoperable (1663); Aspiration Issue (2883)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/27/2018
Event Type  malfunction  
Manufacturer Narrative
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A nurse reported the tubing (hose) did not work as it was not aspirating during a procedure.No known harm to the patient.Additional information was requested; however, none has been received to date.
 
Manufacturer Narrative
There have been no additional complaints reported against the finish goods lot and the device history record shows the product was released per specifications.The returned sample was visually and functionally tested and passed testing, no aspiration issues were found during testing.The root cause of the customer's complaint could not be established; the returned sample met specifications.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined; however potential root causes are listed.After a thorough investigation of this complaint, it has been determined that no action will be taken at this time as the sample met specifications.Quality assurance will continue to monitor customer complaints via the complaint review meetings, and will take action for any future occurrences as is deemed necessary.Consumables manufacturing management has been made aware of this complaint through the consumer affairs review meeting.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7584261
MDR Text Key110925649
Report Number2028159-2018-01193
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeCI
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750833
Other Device ID Number380657508334
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/15/2018
Initial Date FDA Received06/08/2018
Supplement Dates Manufacturer Received08/14/2018
Supplement Dates FDA Received09/07/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-