• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/14/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.\ device not returned.
 
Event Description
Livanova deutschland received a report that an s5 roller pump lost connection to the mast roller pump during a procedure.The user reset the stop link and rebooted the device and the issue did not recur.There was no report of patient injury.
 
Manufacturer Narrative
Through follow-up communication, livanova (b)(4) learned that the follower pump was being used for selective cerebral perfusion, and that the master-follower configuration of the pumps was lost.Livanova (b)(4) was provided the log files of the pumps.A read out analysis was performed at livanova (b)(4), which revealed that there were no events about a log out of the master pump.This implies that the connection between master and follower pump was never taken out.The reported issue could not be confirmed.As the issue could not be confirmed, a root cause was not identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
MDR Report Key7589289
MDR Text Key110707197
Report Number9611109-2018-01032
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 07/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/14/2018
Initial Date FDA Received06/11/2018
Supplement Dates Manufacturer Received06/13/2018
Supplement Dates FDA Received07/31/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-