• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE Back to Search Results
Model Number 3058
Device Problems Electromagnetic Interference (1194); Device Operates Differently Than Expected (2913); Improper Device Output (2953)
Patient Problem Pain (1994)
Event Date 06/08/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp) regarding a patient implanted with an implantable neurostimulator (ins) for gastrointestinal/pelvic floor.It was reported the patient had shooting pain down their right side.The hcp stated the patient did not turn the right-side ins off before going into the mri.The hcp noted after exiting the mri when the shooting pain was noticed, the hcp stated the patient was having difficulty turning the ins off.The patient alleged the right-side ins, the ins that was left on for some duration during the mri, voltage had changed without the patient adjusting or changing the program.The hcp stated the stimulation changed from 1.4 to 2.0.The hcp stated the ins on the left side was 0.0 v.There were no further complications that have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM II
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7589701
MDR Text Key110710277
Report Number3004209178-2018-13138
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00613994913654
UDI-Public00613994913654
Combination Product (y/n)N
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2014
Device Model Number3058
Device Catalogue Number3058
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/08/2018
Initial Date FDA Received06/11/2018
Date Device Manufactured12/13/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
-
-