It was reported that a patient underwent a procedure with a smartablate¿ irrigation tubing set and when flushing was conducted, it was noticed that there was something like dust in the lumen of the tubing.The tubing was replaced to complete the procedure.No patient consequences were reported.The response received confirmed that the material did not have movement; however, the material was confirmed to be inside the tubing.Pictures of the foreign material are not available.The foreign material found inside the tubing was assessed as a reportable malfunction.
|
The biosense webster inc.Failure analysis lab received the device for evaluation on 6/22/2018.The device history record and expiration date have been provided on 7/17/2018.Therefore, expiration date has been populated.Investigation summary: it was reported that a patient underwent a procedure with a smartablate¿ irrigation tubing set and when flushing was conducted, it was noticed that there was something like dust in the lumen of the tubing.The tubing was replaced to complete the procedure.No patient consequences were reported.The response received confirmed that the material did not have movement; however, the material was confirmed to be inside the tubing.Complaint product was inspected and it was found in normal condition.No foreign material found.Irrigation test was performed and the tubing passed specifications.No microbubbles were found in the tubing after flushing.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer's ref.No: (b)(4).
|