OBERDORF SYNTHES PRODUKTIONS GMBH 2.7/3.5MM TI VA-LCP MEDL DSTL HUM PL/1H/LT/69MM-SHORT-STER; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 04.117.501S
Device Problem Material Fragmentation (1261)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Patient information is unknown.It is unknown when metal fragments were generated.Additional procode: hwc.It is unknown if or when device was explanted.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter email address is unknown.Part: 04.117.501s; lot: l557541; manufacturing location: (b)(4); release to warehouse date: september 13, 2017; expiry date: september 01, 2027; no nonconformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes europe reports an event in (b)(6) as follows: it was reported that the open reduction internal fixation (orif) was applied to distal humeral fractures on (b)(6) 2018.String-like metal fragments were found when screws were fixed to plates as follows: five screws = 2x va locking screws; 3x va locking screws.Two plates = 1x va-lcp dhp plate, posterolateral: one screw was inserted; 1x va-lcp dhp plate, medial: four screws were inserted.It is unknown which screws were fixed to which plates.It is unknown from where the fragments came.The fragments were removed.The screw was variable angle inserted, and the drill sleeve was secured to the plate.It was confirmed by x-rays that directions of the screw insertions and drill holes matched.Surgery was completed without further incident.There was no adverse consequence to the patient.This report is for a 2.7/3.5mm ti va-lcp medl dstl hum pl/1h/lt/69mm-short-ster.This is report 5 of 7 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7/3.5MM TI VA-LCP MEDL DSTL HUM PL/1H/LT/69MM-SHORT-STER
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK MEZZOVICO (CH)
via cavazz 5
mezzovico 6805
SZ   6805
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7601046
MDR Text Key111203501
Report Number8030965-2018-54332
Device Sequence Number1
Product Code HRS
UDI-Device Identifier07611819504974
UDI-Public(01)07611819504974
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K120070
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04.117.501S
Device Lot NumberL557541
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/21/2018
Initial Date FDA Received06/14/2018
Date Device Manufactured09/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-