MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number BI-700-00027-120
Device Problems Computer Software Problem (1112); Communication or Transmission Problem (2896); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/20/2017
Event Type  malfunction  
Manufacturer Narrative
A phone representative answered the sites question regarding the imaging system.It was reported that when the manufacturer representative (rep) looked at the ip settings they realized the ip setting for the navigation system on the imaging system were incorrect.Once the issue was fixed the system worked as intended.
 
Event Description
Medtronic received information regarding an imaging system being used in a cranial resection.It was reported that the imaging system and the navigation system were not communicating even though they were connected to each other.During trouble shooting the site found that the ip setting for the navigation system were incorrect.The procedure was delayed 10 minutes.No impact on patient outcome.
 
Manufacturer Narrative
Correction to the initial reporter.
 
Manufacturer Narrative
No system checkout was performed and no analysis was completed on the system.Device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
stephanie riley
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7602567
MDR Text Key111678875
Report Number1723170-2018-02783
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/20/2017
Initial Date FDA Received06/14/2018
Supplement Dates Manufacturer Received09/17/2018
09/17/2018
Supplement Dates FDA Received09/17/2018
09/17/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/23/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-