• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORAX MEDICAL, INC. FENIX CONTINENCE RESTORATION SYSTEM; IMPLANTED FECAL INCONTINENCE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORAX MEDICAL, INC. FENIX CONTINENCE RESTORATION SYSTEM; IMPLANTED FECAL INCONTINENCE DEVICE Back to Search Results
Model Number FS18
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Erosion (1750); Pain (1994)
Event Date 04/25/2018
Event Type  Injury  
Event Description
Following a surgical procedure for reinforcement of the anal sphincter due to fecal incontinence, a patient was found to have one fenix bead eroded through the anal canal leading to fenix device explant.The fenix device was used as part of the surgical procedure.Uneventful surgical procedure and device implant on (b)(6) 2014.Clinical examination on (b)(6) 2018 revealed "right anterior skin ulceration which showed part of the magnetic sphincter 1 cm from the anus".The patient reported recent perianal pain with "no fever or suspicious oozing".Uneventful device explant (b)(6) 2018 due to 1 fenix device bead found eroded through the anterior right anal canal.The entire fenix device was removed through the original incision.The device was intact at the time of explant.It was observed that "there is a fibrous perianal tunnel, but the anus is still gaping.The rectal mucosa is visible, but does not go past the marginal plane, even under traction".It was also observed that there was no infection at the exposure site.The patient was reported as "doing as well [as] its possible" after as of (b)(6) 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FENIX CONTINENCE RESTORATION SYSTEM
Type of Device
IMPLANTED FECAL INCONTINENCE DEVICE
Manufacturer (Section D)
TORAX MEDICAL, INC.
4188 lexington avenue n
shoreview MN 55126
Manufacturer (Section G)
TORAX MEDICAL, INC.
4188 lexington avenue north
shoreview MN 55126
Manufacturer Contact
tressa lauer
4188 lexington avenue north
shoreview, MN 55126
6513618900
MDR Report Key7603232
MDR Text Key111115122
Report Number3008766073-2018-00098
Device Sequence Number1
Product Code PMH
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
H130006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 05/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date07/11/2016
Device Model NumberFS18
Device Lot Number3740
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/16/2018
Initial Date FDA Received06/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/11/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Disability;
Patient Age83 YR
-
-