Model Number 25-092-42-91
Device Problem Break (1069)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 05/24/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported a screw broke.A portion of the screw remains in the patient.
 
Manufacturer Narrative
The stereo microscope investigation revealed tensile cracks.Further observation determined there were no indications of material or manufacturing defects.The review of the product device history records was not possible due to no lot number being identified.The investigation results conclude that the root cause for breakages were due to structural failure of the device due to mechanical overload.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Type of Device
SCREW
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau 78570
GM  78570
Manufacturer (Section G)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau 78570
GM   78570
Manufacturer Contact
jennifer damato
p.o. box 16369
jacksonville, FL 32245
9046417746
MDR Report Key7607352
MDR Text Key111266760
Report Number9610905-2018-00087
Device Sequence Number1
Product Code DZL
UDI-Device Identifier00888118037268
UDI-Public(01)00888118037268
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083432
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model Number25-092-42-91
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date05/24/2018
Event Location Hospital
Date Report to Manufacturer05/24/2018
Initial Date Manufacturer Received 05/24/2018
Initial Date FDA Received06/15/2018
Supplement Dates Manufacturer Received06/29/2018
Supplement Dates FDA Received07/25/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age37 YR
-
-